 The
slides to be stained for
CD10 contained four small
lymphocytic lymphomas/chronic
lymphocytic
leukaemias (specimens
1-4), two normal kidneys
(specimens 5-6),
a
colon carcinoma
(specimen 7),
three Hodgkin lymphomas
LP (specimens
8-10), and three follicular
lymphomas (specimens
11-13).
43 laboratories
submitted stainings. All laboratories used
mAb 56C6 from either Novocastra (most), Neomarkers or Ventana. All
used
HIER as
pre-treatment.
At the assessment, 14
laboratories (33%) achieved optimal staining, 13 (30%) acceptable, 11 (26%) borderline, and 5 (12%) poor.
The basis for an optimal result was an intense
and distinct staining of cell membranes in cells expected to
stain.
Mandatory for an optimal CD10 staining reaction was an efficient HIER protocol combined
with a sensitive visualization system
and a method to avoid endogenous biotin reaction.
All laboratories achieving an optimal
demonstration of CD10 used HIER in an alkaline buffer
(10/1 mM
Tris/EDTA or Tris/EGTA), and 13 of 14 used a polymer based visualization system
(EnVision), while
one used an
avidin/biotin system with an efficient blocking of endogenous
biotin.
|