|
In
this first run H1 in the new HER-2 ISH module both BRISH and FISH
tests are offered. The BRISH tests are carried out like IHC tests,
i.e., the stains are
evaluated by an assessor group and the corresponding protocols
analysed to identify recommendable probes and methods, as well as
causes of insufficient results. This is not possible for FISH tests,
as FISH stained slides can not be submitted for external assessment.
Common for BRISH and FISH tests in the NordiQC scheme is the scoring
challenge aiming at the laboratories' proficiency in scoring the
HER-2 and chromosome 17 signals. This allows laboratories performing
FISH tests to ensure if their combined staining and scoring is
sufficient. The results of the FISH scorings are shown in Table 2
and Figure 1.
The
material circulated for the
HER-2 ISH assessment run comprised one
normal breast tissue & four breast ductal carcinomas showing the
HER-2 gene/chromosome 17 (HER-2/chr17) ratios as follows:
|
|
HER-2
IHC* |
Dual - SISH** |
FISH*** |
|
|
IHC score |
HER-2/chr17 ratio |
HER-2/chr17 ratio |
|
1. Normal breast
tissue |
0 |
1.1 - 1.2 |
1.1 - 1.3 |
|
2. Breast ductal
carcinoma |
3+ |
3.5 - 4.0 |
4.3 - 5.5 |
|
3. Breast ductal
carcinoma |
1+ |
1.2 - 1.4 |
1.3 - 1.5 |
|
4. Breast ductal
carcinoma |
2+ |
1.5 - 1.8 |
1.8 - 2.2 |
|
5. Breast ductal
carcinoma |
2+ |
1.3 - 1.6 |
1.5 - 1.9 |
*PATHWAY®, Ventana (data from one
reference lab.), **Inform HER-2 Dual SISH kit, Ventana (range of data from two
reference labs.), ***HER2 FISH pharmDX™ Kit, Dako (range of data from two
reference labs.).
All tissues were fixed for 24 - 48 h. in 10 %
neutral buffered formalin (NBF).
HER-2 BRISH, technical assessment
The main criteria for assessing a BRISH HER-2 analysis as technically optimal
was the ability to interpret
and evaluate the HER-2/chr17 ratios in all five tissues.
A staining was assessed as good, if the HER-2/chr17 ratios could be
evaluated in all five tissues, but the interpretation was slightly compromised
to e.g. due to a weak or excessive counterstaining, excessive retrieval or
similar.
A staining was assessed as borderline if one of the tissue cores could not be properly
evaluated e.g. because of too weak signals, no signals and/or a low signal-to-noise ratio.
A staining was assessed as poor if two or more of the tissue cores could not be
properly evaluated.
HER-2 BRISH and FISH interpretation
For both BRISH and FISH the participating laboratories were asked to submit a
scoring sheet with the interpretation of the HER-2/chr17 ratio
for all the five tissues. The NordiQC FISH data was used as reference in order
to evaluate the scoring consensus between the participating laboratories and
NordiQC. A concordant interpretation was seen if the five tissues were evaluated
as listed below:
-
Evaluation of the normal breast tissue no. 1. corresponding a non-amplified status.
-
Evaluation of the breast ductal carcinoma no.
2 corresponding an amplified status
-
Evaluation of the breast ductal carcinoma no.
3 corresponding a non-amplified status.
-
Evaluation of the breast ductal carcinoma no.
4 corresponding an equivocal (borderline) status.
-
Evaluation of the breast ductal carcinoma no.
5 corresponding a non-amplified or equivocal (borderline)status.
Results BRISH
In total 105 laboratories participated in this assessment. 84
laboratories performed BRISH and out of these 68 (81 %) achieved a
sufficient mark. The results are summarized in Table 1.
Table 1.
Systems and
assessment marks
for BRISH HER-2, run B12
|
Two colour HER-2 systems |
N |
Vendor |
Optimal |
Good |
Borderl. |
Poor |
Suff.1 |
Suff.
OPS2 |
INFORM™ HER-2 Dual
ISH
780-4332+780-4331 800-4422 |
57 |
Ventana |
32 |
12 |
7 |
6 |
77 % |
76 % |
|
DuoCISH™ SK109 |
9 |
Dako |
3 |
3 |
2 |
1 |
67 % |
75 % |
DuoCISH™
SK108 + K5331 |
3 |
Dako |
3 |
0 |
0 |
0 |
- |
- |
|
Unknown |
1 |
Unknown |
0 |
1 |
0 |
0 |
- |
- |
|
One colour HER-2 systems |
|
|
|
|
|
|
|
|
INFORM™ HER-2 SISH
780-4332 |
7 |
Ventana |
6 |
1 |
0 |
0 |
100 % |
100 % |
|
ZytoDot ®
C-3003 |
4 |
ZytoVision |
4 |
0 |
0 |
0 |
- |
- |
|
SPOT-Light®
84-0150 |
2 |
Invitrogen |
2 |
0 |
0 |
0 |
- |
- |
|
“In-house” |
1 |
|
0 |
1 |
0 |
0 |
- |
- |
|
Total |
84 |
|
50 |
18 |
9 |
7 |
- |
- |
|
Proportion |
|
|
60 % |
21 % |
11 % |
8 % |
81 % |
|
1) Proportion of sufficient stains. 2) Proportion of sufficient stains with optimal
protocol settings only, see below.
Comments
In this assessment a sufficient demonstration and evaluation of the
HER-2 gene amplification status in all the tissues included in the
multitissue block could be obtained by all the different BRISH systems
used by the laboratories. All the included tissues were fixed in 10
% neutral buffered formalin for 24-48 hours according to the ASCO/CAP
guidelines for the tissue preparation of breast tissue. However the
breast ductal carcinoma, tissue no. 4 was slightly more challenging
than the other tissues regarding the protocol settings as this
tumour was found to be less robust as excessive retrieval typically impaired the morphology
and thus complicating the identification of the BRISH signals. This
pattern was seen for all systems used.
Optimal protocol settings
Two-colour HER-2 systems
For the INFORM™ Dual ISH system, Ventana, an optimal
demonstration for HER-2 BRISH was in brief typically based upon HIER
in Cell Conditioning 2 (CC2) for 24-32 min. at 86-90˚C and
proteolysis in P3 for 8 - 16 min at 36-37˚C. The HER-2 SISH probe
was typically applied for 6 hours at 50-52˚C, while the chr17
probe was applied for 2 hours at 42 - 44˚C.
Using these protocol settings a sufficient result was seen in 76 %
of the submitted protocols (35 out of 46). 11 laboratories used a
protocol with optimal settings, but for unexplained reasons a
complete false negative staining reaction or a staining result with
an excessive background staining e.g. due to silver precipitates was
seen. The remaining insufficient results were characterized by an
impaired morphology. This pattern was typically caused by excessive
retrieval hampering the interpretation as the nuclei were almost
totally digested complicating the identification and interpretation
of the BRISH signals.
For the DuoCISH™ system SK109, Dako, the main protocol
settings giving an optimal result were based on HIER for 10 min in
the pre-treatment buffer at 95 - 98°C and proteolysis in Pepsin for
2-3 min. at 37°C or 7 - 10 min. at room temp. (both reagents
included in the HER2 CISH pharmDX kit SK109). The HER-2 and the chr17 probes were applied for 14 – 20 hours at 45˚C and visualized by
the DuoCISH™ kit SK109, Dako.
Using these protocol settings a sufficient result was seen in 75 %
of the submitted protocols (6 out of 8). In the insufficient results
typically a too weak or false negative staining for the HER-2
signals in both the neoplastic cells and in the normal stromal cells
was seen. This observation might be related to a too low sensitivity
of the reagents used for the immunohistochemical demonstration of
the HER-2 genes. In this context it is of
utmost importance that the Red chromogene used for the visualization
of the HER-2 genes in the DuoCISH™ system is prepared immediately
before use and that Pepsin is stored at appropriate conditions in
order to maintain the proteolytic capacity.
One-colour HER-2 systems
For the INFORM™ HER-2 SISH, Ventana, an optimal result
typically was based upon HIER in CC2 or Reaction buffer (RB) for
24-32 min. at 90-95˚C and proteolysis in P3 for 4 - 8 min at 37˚C.
The HER-2 SISH probe was typically applied for 6 hours at 50-52˚C.
Using these protocol settings a sufficient result was seen in 100 %
of the submitted protocols (7 out of 7).
For the ZytoDot® CISH system C-3003, ZytoVision, an optimal
result typically was obtained by using proteolysis in Pepsin for 3
min at room temp, HIER in EDTA for 10-15 min. at 98˚C, hybridization
at 37˚C for 14-16 hours and visualized by the ZytoVision detection
kit C-3003.
For the SPOT-Light® CISH system 84-0150, Invitrogen, an
optimal result was obtained by using proteolysis in Pepsin for 5-7
min. at room temp, HIER in Tris-EDTA for 15 min. at 98˚C,
hybridization at 37˚C for 14 hours and visualized by the Invitrogen
detection kit 84-0150.
HER-2 interpretation and scoring consensus
Both the laboratories performing BRISH and FISH were requested to
send in their own interpretation on the stained sections, which was
completed by 98 out of the 105 laboratories participating in this
run. The participants evaluations were compared to the HER2 FISH
amplification status generated by the NordiQC reference
laboratories. The overall results are summarized in table 2.
No significant
difference in concordance rates was seen between the participants
using FISH or BRISH.
Table 2: Interpretation and scoring consensus between the
participants and the NordiQC FISH data
|
|
FISH HER-2/chr17 ratio |
HER2
amplification status |
Consensus evaluation |
|
1.
Normal breast tissue |
1.1 - 1.3 |
Non-amplified |
99 % |
|
2.
Breast ductal carcinoma |
4.3 - 5.5 |
Amplified |
100 % |
|
3.
Breast ductal carcinoma |
1.3 - 1.5 |
Non-amplified |
71 % |
|
4.
Breast ductal carcinoma |
1.8 -
2.2 |
Equivocal (Borderline) |
43 % |
|
5. Breast ductal carcinoma |
1.5 - 1.9 |
Non-amplified / Equivocal |
90 % |
A high consensus rate (99-100%) between the participants and NordiQC
regarding the determination of the HER-2 amplification status was
seen for the normal breast tissue no.1 and the highly amplified
breast carcinoma, tissue no. 2, whereas a lower consensus rate was
seen in the breast carcinomas. tissues no. 3 and 5 (71–90%). For the
breast carcinoma, tissue no. 4 with a HER-2/chr17 ratio in the range
of 1.8 – 2.2 based on FISH (thus to be categorized as
equivocal/borderline) a low
consensus rate (43 %) was seen. However, at stated above, even one
of the reference labs. obtained a score of 1.5 with dual-SISH (i.e.,
no consensus).
This emphasizes the need both to focus on the technical set-up of the ISH procedure and to establish
a better standardization to support and harmonize the interpretation of
the ISH results. The recommendations for HER-2 testing in the UK1
as well as digital image analysis may aid in more standardized and
reproducible results.
1. Bartlett JM, Starczynski J, Atkey N, Kay E, O'Grady A, Gandy M,
Ibrahim M, Jasani B, Ellis IO, Pinder SE, Walker RA. HER2 testing in
the UK: recommendations for breast and gastric in-situ hybridisation
methods. J Clin Pathol. 2011 Aug;64(8):649-53. Epub 2011 Jun 20.
PubMed PMID: 21690244.
This was the 7th NordiQC assessment of HER-2 BRISH/ISH. As
seen in table 3, similar pass rates have been obtained in the
last two assessments.
Table 3:
Proportion of sufficient results for HER-2 BRISH in the seven NordiQC runs
performed
Conclusion
In this assessment an optimal demonstration of HER-2 BRISH could be
obtained by the two commercially available two-colour HER-2 systems
INFORM™ HER-2 Dual ISH, Ventana and DuoCISH™,Dako.
Also the single-colour HER-2 systems, INFORM™ HER-2 SISH, Ventana,
ZytoDot®, ZytoVision and SPOT-Light®, Invitrogen could be used to
obtain an optimal demonstration.
For an optimal performance the retrieval settings – HIER +
proteolysis – must be carefully balanced to provide both a high sensitivity and
a preserved morphology. Attention must also be
addressed to the interpretation as a high inter-observer variation
was seen. |