Home  ■  Protocols

Recommended protocol Run 33 2011
Odense C, DK

CK-LMW (EP17)

Protocol provided by

 
  laboratory Institute of Pathology
  hospital/institution Odense Universityhospital,    Odense C, Denmark
  e-mail address ole.nielsen@ouh.regionsyddanmark.dk
Primary antibody  
  clone EP17
  producer Epitomics
  product no., lot no. AC-0007, 5555
  dilution 1:100
  diluent buffer and additive(s) Ab diluent S2022, Dako
  incubation time/temp. 32 min./36°C
Epitope retrieval, HIER  
  device BenchMark Ultra
  buffer type, pH Cell Conditioning 1, pH 8.5
  warm-up/heating max/resting time 8 min./64 min./8 min.
  maximum heating temp. 99°C
Visualization system  
  method 3-step multimer conjugate
  producer, product no. Ventana, 760-500
  incubation time/temp. 8 min./36°C x 3
Chromogen  
  type DAB
  producer, product no. Ventana, 760-500
  incubation time/temp. 8 min./36°C
  enhancement, type CuSO4
Immunostainer  
  type BenchMark Ultra, Ventana
     
Abbreviations  
  HIER: heat induced epitope retrieval, MWO: microwave oven, NS: not stated, RT: room temperature
Disclaimer  
  NordiQC makes every attempt to provide accurate and up-to-date information, yet NordiQC does not make any claim or warranty regarding the accuracy of the provided information nor does it represent that the contents of the web site and protocols reflect the most recent developments in immunohistochemistry at any point in time.
 
SN/CJ

Last update 29-11-2011