The
slide to be stained for PR
comprised the following five tissues:
|
No. |
Tissue |
PR-positivity* |
PR-intensity* |
|
1. |
Uterine cervix |
80-90 % |
Moderate to strong |
|
2. |
Breast ductal carcinoma |
0 % |
Negative |
|
3. |
Breast ductal carcinoma |
40 - 60 % |
Weak to moderate |
|
4. |
Breast ductal carcinoma |
30 - 50 % |
Weak to moderate |
|
5. |
Breast ductal carcinoma |
80 - 100 % |
Moderate to strong |
*PR-status and staining pattern as
characterized by NordiQC reference laboratories using the mAb
clone
16 (Leica/Novocastra) and the rmAb clone 1E2 (Ventana).
All tissues were fixed in 10% neutral buffered formalin for 24 – 48
hours and processed according to Yaziji et al. (1).
Criteria for assessing a PR staining as optimal included:
-
A moderate to strong, distinct nuclear staining reaction of both the
columnar and basal squamous epithelial cells and most of the stromal
cells (with the exception of endothelial cells and lymphoid cells)
in the uterine cervix.
-
An at least weak to moderate distinct nuclear staining reaction in
the appropriate proportion of the neoplastic cells in the breast
ductal carcinomas no. 3, 4 & 5.
-
No nuclear staining reaction of the neoplastic cells in the breast
carcinoma no. 2 and no more than a weak cytoplasmic reaction in
cells with a strong nuclear staining.
For the mAb clones PR 636 and 1A4 a cytoplasmic staining in the
columnar epithelial cells and the intermediate squamous epithelial
cells, respectively, was accepted.
A staining was classified as good if ≥ 10 % of the neoplastic cells
in the breast ductal carcinomas no. 3, 4 & 5 showed an at least weak
nuclear staining reaction but below the reference ranges.
A staining was assessed as borderline if ≥ 1 % and < 10 % of the
neoplastic cells showed a nuclear staining reaction in one or more
of the breast ductal carcinomas no. 3, 4 & 5.
A staining was assessed as poor if < 1% of the neoplastic cells
showed a nuclear staining reaction in one or more of the breast
ductal carcinomas no. 3, 4 and 5 or a false positive nuclear
staining reaction was seen in the breast ductal carcinoma no. 2.
1. Yaziji H, Taylor CR, Goldstein NS, Dabbs DJ, Hammond EH, Hewlett
B, Floyd AD, Barry TS, Martin AW, Badve S, Baehner F, Cartun RW,
Eisen RN, Swanson PE, Hewitt SM, Vyberg M, Hicks DG; Members of the
Standardization Ad-Hoc Consensus
Committee. Consensus recommendations on estrogen receptor testing in
breast cancer by immunohistochemistry. Appl Immunohistochem Mol
Morphol. 2008 Dec;16(6):513-20. PubMed PMID: 18931614.
224 laboratories participated in this assessment. 68 % achieved a
sufficient mark. In table 1 the antibodies (Abs) used and marks are
summarized.
Table 1. Abs and
assessment marks
for PR, run B12
|
Concentrated Abs |
N |
Vendor |
Optimal |
Good |
Borderl. |
Poor |
Suff.1 |
Suff. OPS2 |
|
mAb clone
1A6 |
3
1 |
Leica/Novocastra
Vector |
0 |
4 |
0 |
0 |
- |
- |
|
mAb clone
16 |
21
1
1 |
Leica/Novocastra
Monosan
Vector |
10 |
5 |
6 |
2 |
65 % |
91 % |
|
mAb clone cocktail
16 + SAN27 |
2 |
Leica/Novocastra |
0 |
1 |
1 |
0 |
- |
- |
|
mAb clone cocktail
hPRa2+hPRa3 |
1 |
Thermo/NeoMarkers |
0 |
0 |
0 |
1 |
- |
- |
mAb clone
PR 1294 |
3 |
Dako |
0 |
2 |
1 |
0 |
- |
- |
mAb clone
PR 636 |
52 |
Dako |
13 |
14 |
8 |
17 |
52 % |
69 % |
|
rmAb clone
RBT22 |
2 |
Bio SB |
0 |
0 |
1 |
1 |
- |
- |
|
rmAb clone
SP2 |
4 |
Thermo/NeoMarkers |
1 |
0 |
0 |
3 |
- |
- |
|
rmAb clone
SP42 |
1 |
Zytomed |
0 |
0 |
1 |
0 |
- |
- |
|
rmAb clone
Y85 |
2
1 |
Cell Marque
Master Diagnostica |
1 |
0 |
2 |
0 |
- |
- |
|
Ready-To-Use Abs |
|
|
|
|
|
|
|
|
|
mAb clone
16,
PA0312 |
5 |
Leica/Novocastra |
1 |
4 |
0 |
0 |
100 % |
100 % |
|
mAb clone
16, PM424 |
1 |
Biocare |
0 |
1 |
0 |
0 |
- |
- |
|
mAb clone
PR 1294, SK310/K4071 |
5 |
Dako |
0 |
0 |
2 |
3 |
0 % |
- |
|
mAb clone PR 636, IS/IR068 |
33 |
Dako |
15 |
10 |
5 |
3 |
76 % |
85 % |
|
mAb clone PR 636, N1630 |
1 |
Dako |
0 |
0 |
1 |
0 |
- |
- |
|
rmAb clone 1E2,
790-223/4296 |
83 |
Ventana |
31 |
38 |
11 |
3 |
83 % |
85 % |
|
rmAb clone SP2,
ZA0255 |
1 |
Zhongshan |
0 |
1 |
0 |
0 |
- |
- |
|
Total |
224 |
|
72 |
80 |
39 |
33 |
- |
|
|
Proportion |
|
|
32 % |
36 % |
17 % |
15 % |
68 % |
|
1) Proportion of sufficient stains
(optimal or good), 2) Proportion of sufficient stains with optimal
protocol settings only, see below.
The following central protocol parameters were used to obtain an
optimal staining:
Concentrated Abs
mAb clone 16: The protocols giving an optimal result were all based
on heat induced epitope retrieval (HIER) using either Tris-EDTA/EGTA
pH 9 (1/1)*, Target Retrieval Solution (TRS) pH 9 (3-in-1) (Dako)
(2/2)*, Bond Epitope Retrieval Solution 2 (BERS2; Bond, Leica)
(5/6)*, BERS1 (Bond, Leica) (1/1)* or EDTA/EGTA pH 8 (1/1)* as the
retrieval buffer. The mAb was typically diluted in the range of
1:200 – 1:1,600 depending on the total sensitivity of the protocol
employed. Using these protocol settings 10 out of 11 (91 %)
laboratories produced a sufficient staining (optimal or good).
*(number of optimal results/number of laboratories using this
buffer)
mAb clone PR 636: The protocols giving an optimal result were all
based on HIER using either Tris-EDTA/EGTA pH 9 (4/11), TRS pH 9
(Dako) (2/11), TRS pH 9 (3-in-1,Dako) (6/11) or BERS2 (Bond, Leica)
(1/4) as the retrieval buffer. The mAb was typically diluted in the
range of 1:100 – 1:500 depending on the total sensitivity of the
protocol employed. Using these protocol settings 20 out of 29 (69 %)
laboratories produced a sufficient staining.
rmAb clone SP2: The protocol giving an optimal result was based HIER
using Tris-EDTA/EGTA pH 9 (1/2) as the retrieval buffer. The rmAb
was diluted 1:200.
rmAb clone Y85: The protocol giving an optimal result was based on
HIER using EDTA/EGTA pH 8 (1/1) as the retrieval buffer. The rmAb
was diluted 1:100.
Ready-To-Use Abs
mAb, clone 16 (prod.no. PA0312, Leica/Novocastra). The protocol
giving an optimal result was based on HIER in BERS2 (Bond, Leica)
for 25 min, an incubation time of 15 min in the primary Ab and
Refine (DS9800) as the detection system. Using similar protocol
settings 3 out of 3 produced a sufficient staining.
mAb clone PR 636 (prod. no. IS/IR068, Dako): The protocols giving
an optimal result were all based on HIER in PT-Link using TRS pH 9
or TRS pH 9 (3-in-1) (heating time 10-30 min at 95-97°C) , an
incubation time of 15-30 min in the primary Ab and EnVision Flex
(K8000) or Flex+ (K8002) as the detection system. Using these
protocol settings 23 out of 27 (85 %) laboratories produced a
sufficient staining.
rmAb clone 1E2 (prod. no. 790-2223/4296, Ventana): The protocols
giving an optimal result were typically based on HIER using mild or
standard Cell Conditioning 1, an incubation time of 20-48 min in the
primary Ab and iView (760-091) or UltraView (760-500) as the
detection system. Using these protocol settings 56 out of 66 (85 %)
laboratories produced a sufficient staining (optimal or good).
The most frequent causes of insufficient stains were:
- Too low concentration of the primary Ab
- Less successful primary Ab
- Insufficient epitope retrieval (too
short efficient HIER time and/or use of a non-alkaline HIER buffer).
In this assessment and in concordance with the previous PR
assessments in NordiQC, the insufficient results were mainly due to
a too weak or completely false negative staining. This pattern was
seen in 89 % of the insufficient staining results (65 out of 72
stains). Virtually all the laboratories could demonstrate PR in the
ductal breast carcinoma no. 5 with 80-100% positivity and a strong
nuclear staining intensity (as established by the NordiQC reference
laboratories) whereas the prevalent feature of the insufficient
staining was a too weak (< 10% positivity) or entirely false
negative staining of the ductal breast carcinoma no. 3 and 4 (with
30-60% positivity and a weak to moderate nuclear staining intensity
expected). The insufficient staining reactions were typically caused
by a too low concentration of the primary Ab and/or insufficient
HIER, but also when using less successful Abs with an apparently low
PR affinity. However it also has to be mentioned that a high
proportion of stains assessed as insufficient (18 out of 72
protocols) seemed to be based on optimal protocol settings,
including a correct titre of an otherwise successful primary Ab,
HIER settings and detection kit. Thus, the combination of several
suboptimal protocol settings might have caused a reduced
sensitivity, e.g., a low concentration of the primary Ab, HIER in a
non-alkaline buffer and a detection system with only a moderate
sensitivity such as a two-step polymer based kit. For the
laboratories using a non-alkaline buffer for HIER the overall pass
rate was 34 % (13 out of 38), compared to a pass rate of 74 % when
HIER was performed with an alkaline buffer. The same pattern was
seen when the pass rates were related to the detection system: With a
moderately sensitive detection system,
such as a 2-step polymer/multimer based detection system or an
avidin-biotin based system, an overall pass rate of 62 % was seen
(105 out of 169) compared to 86 % (47 out of 55) for a 3-step
polymer/multimer based system.
All 4 stains based on the mAb
clone PR 636 and performed on the BenchMark platform (Ventana) were
assessed as insufficient, despite of protocol settings similar to the
settings giving a sufficient result on other IHC platforms, e.g. the
Dako Autostainer.
In 6 % of the insufficient results (4 out of 72 stains) a
combination of a too weak staining and impaired morphology was seen.
This pattern was typically seen by the use of excessive HIER in
combination with a protocol giving a too low sensitivity.
In the remaining 3 insufficient stains (5 %) a false positive
staining in the PR negative ductal carcinoma no. 2 was seen. When a
false positive reaction was observed in the PR negative tumour also
a false positive nuclear reaction was seen in scattered
non-epithelial cells as lymphocytes and endothelial cells. The false
positive staining was seen for the rmAb clone SP2 (2 out of 5
protocols) and the mAb clone cocktail hPRa2+hPRa3 (1 out of 1
protocol). A false positive nuclear reaction was also observed and
described in the previous assessments run B6 and B9 and most likely
caused by a combination of inadequate washing in buffer and too high
concentration of the primary Ab.
For both the mAb clones 16 and PR 636 the Ready-To-use systems from
Leica/Novocastra and Dako, respectively, the obtained pass rates
were higher than the pass rates obtained for the same clones used
with an in-house validated assay.
As observed in the previous assessments for PR, the uterine cervix
seems to be an appropriate control for the evaluation of the
sensitivity of the PR staining: With an optimal protocol almost all
the columnar epithelial cells, the basal squamous epithelial cells
and most of the stromal cells must show a strong and distinct
nuclear staining with only a minimal cytoplasmic reaction. Virtually
all laboratories obtaining this staining pattern were assessed as
sufficient. However, differences regarding the reaction pattern are
seen depending on the Ab selected. When using the mAb clone 1A6, the
basal squamous epithelial cells are negative and a cytoplasmic
reaction is seen in the intermediate and superficial squamous
epithelial cells, while the clone PR 636 gives an intense
cytoplasmic reaction in the columnar epithelial cells.
This was the 6th assessment of PR in NordiQC. A decrease in the
proportion of sufficient results is seen in the two latest runs for
PR as shown in table 2:
Table 2:
Proportion of sufficient results for PR in the six NordiQC runs
performed
In the last two assessments for PR many new laboratories have
enrolled in the NordiQC programme. In the current run a slightly
higher pass rate was observed for the laboratories participating in
all three runs from B6 to B12 than for the laboratories
participating for the first time in either run B9 or B12:
For the laboratories participating for the first time in run B9 or
B12 the pass rate was 64 % (86 out of 134), whereas the pass rate
was 73 % (66 out of 90 laboratories) for the laboratories
participating in all three runs from B6 to B12.
Conclusion
The mAb clones 16, PR 636 and the rmAb clone 1E2 are all well
performing and recommendable Abs for PR. In general the Ready-To-Use
systems for these three Abs showed a superior pass rate compared to
in-house validated assays for PR. Efficient HIER (preferably in an
alkaline buffer) is mandatory for an optimal performance and must be
carried out to provide an optimal balance between sensitivity and
morphology. Uterine cervix is an appropriate control, in which
almost all the columnar epithelial cells, the basal squamous
epithelial cells and most of the stromal cells must show a strong
and distinct nuclear staining reaction with only a minimal
cytoplasmic staining. |