The material circulated for
HER-2 BRISH (Brightfield In Situ Hybridization) assessment run 10 was identical
to the material used in run B9, 2010 and comprised normal breast
tissue & 5 breast ductal carcinomas showing HER-2 gene/chromosome 17
(HER-2/chr17) ratios as follows:
|
|
HER-2/chr17 ratio |
|
|
Duo-CISH* |
Dual - SISH** |
FISH*** |
|
1. Normal breast
tissue1 |
1.1 |
1.0 |
1.1 |
|
2. Breast ductal
carcinoma |
> 6.0 |
> 6.0 |
> 6.0 |
|
3. Breast ductal
carcinoma |
1.4 |
1.3 |
1.3 |
|
4. Breast ductal
carcinoma |
2.5 |
2.6 |
2.3 |
|
5. Breast ductal
carcinoma |
2.2 |
2.7 |
2.3 |
|
6. Breast ductal
carcinoma |
2.4 |
2.8 |
2.6 |
* HER-2 DuoCISH™ kit, Dako (data from one
ref. lab.), ** HER-2 Dual SISH kit, Ventana (average of data
from two ref. labs.),
*** HER2 FISH pharmDX™ Kit, Dako (average
of data from three tests perfomed in one ref. lab.).
All carcinomas were fixed for 24 h in 10 % neutral buffered formalin
(NBF), except for the carcinoma in tissue core nos. 5 and 6, the
same tumour fixed for 48 and 72 h. respectively.
Criteria for assessing a BRISH HER-2 analysis as optimal included:
-
Staining of the normal breast tissue and the ductal carcinoma no. 3
corresponding a non-amplified status.
-
Staining of breast ductal carcinomas no. 2, 4, 5 and 6 corresponding
an amplified status
-
Staining with preserved morphological details and a minimal
background reaction.
A staining was assessed as good, if the above mentioned criteria
were fulfilled for the normal breast tissue (no. 1) and the three
ductal carcinomas fixed for 24 h (no. 2-4), but not fulfilled for
the carcinoma no. 5 and 6 fixed for
48 and 72 h. respectively in NBF.
Even though carcinoma no. 6 was overfixed according to the ASCO/CAP guidelines,
it was from a technical perspective important to see if the labs could carry out a
successful CISH/SISH procedure also for this tumour in spite of
prolonged fixation. The tumour had successfully been evaluated
by FISH for HER-2 in the NordiQC reference laboratories.
A staining was assessed as borderline if one of the other carcinomas
could not be properly evaluated due to a too weak signal or a low
signal-to-noise ratio.
A staining was assessed as poor in case that more of the other
carcinomas could not be properly evaluated.
Results
57 laboratories participated in this assessment. 41 (72 %) achieved
a sufficient mark. The results are summarized in Table 1.
Table 1.
Abs and
assessment marks
for CISH/SISH HER-2, run B10
|
Two colour HER-2 systems |
N |
Vendor |
Optimal |
Good |
Borderl. |
Poor |
Suff.1 |
|
INFORM™ HER-2 Dual SISH |
27 |
Ventana |
10 |
11 |
3 |
3 |
77 % |
|
DuoCISH™ |
11 |
Dako |
3 |
4 |
3 |
1 |
64 % |
|
ZytoDot® 2 C |
2 |
ZytoVision |
0 |
2 |
0 |
0 |
- |
|
One colour HER-2 systems |
|
|
|
|
|
|
|
|
SPOT-Light® |
4 |
Invitrogen |
0 |
3 |
1 |
0 |
- |
|
ZytoDot ® |
7 |
ZytoVision |
0 |
3 |
1 |
3 |
43 % |
|
INFORM™ HER-2 SISH |
4 |
Ventana |
2 |
1 |
0 |
1 |
- |
|
“In-house” |
2 |
|
1 |
1 |
0 |
0 |
- |
|
Total |
57 |
|
16 |
25 |
8 |
8 |
- |
|
Proportion |
|
|
28 % |
44 % |
14 % |
14 % |
72 % |
1) Proportion of sufficient stains
(optimal or good), 2) Proportion of sufficient stains with optimal
protocol settings only, see below.
Comments
In this assessment and in accordance with the previous run B9 for
HER-2 BRISH, the two most widely BRISH systems used, the INFORM™
Dual SISH system, Ventana, and the DuoCISH™, Dako could both be used
to obtain an optimal demonstration of the HER-2 gene
amplification status in the tissues included in the multitissue block. In
this assessment also the single colour HER-2 SISH system, Ventana
could be used to obtain an optimal demonstration of HER-2 BRISH.
The most widely used protocol for the INFORM™ Dual SISH system,
Ventana giving an optimal demonstration for HER-2 BRISH was in brief
based upon HIER in Cell Conditioning reaction buffer (CCrb) for 40 –
48 min. at 90˚C and proteolysis in P3 for 8 - 12 min. The HER-2 SISH
probe was typically applied for 6 hours at 50-52˚C, while the Chr.
17 probe was applied for 2 hours at 42 - 44˚C.
For the DuoCISH™ system, Dako, the main protocol settings were based
on HIER for 10 min in the pre-treatment buffer at 95 - 99°C and
proteolysis for 2 min. in Pepsin at 37°C (both reagents included in
the FISH pharmDX kit K5331, Dako). The HER-2 and the Chromosome 17
probe (K5331, Dako) was applied for 14 – 20 hours at 45˚C and
visualized by the DuoCISH™ kit SK108, Dako.
The prevalent feature of an insufficient result was typically a
general too weak or completely negative reaction of the HER-2
signals in both the neoplastic cells and in the normal stromal
cells. This feature can be related to many factors as e.g.
insufficient retrieval or too low sensitivity of the detection
system applied. For the INFORM™ Dual SISH, Ventana the insufficient
retrieval most commonly was noticed as short HIER in CCrb for 20 –
28 min or HIER in CC1. In a few laboratories an insufficient result
was obtained despite optimal protocol settings was applied. At
present NordiQC can not identify any reason for this.
For the DuoCISH™ system, Dako the insufficient retrieval seemed to
be related to insufficient proteolysis in Pepsin – too short time
and/or reduced enzymatic capacity of the applied Pepsin. Pepsin is a
relative fragile enzyme and rapidly deteriorates if stored at room
temperature. Pepsin should always be stored at 2 - 6°C and kept on
ice when taken out of the refrigerator to secure optimal storage
conditions.
In a few laboratories it was also observed that the insufficient
result mainly was related to the demonstration of the HER-2 signals,
whereas the chr17 signals were distinctively demonstrated. This
observation might be related to a too low sensitivity of the
reagents used for the immunohistochemical demonstration of the HER-2
genes. In this context it has to be stressed that it is of utmost
importance that the Red chromogene used for the visualization of the
HER-2 genes in the DuoCISH™ system is prepared immediately before
use.
It was also seen that excessive retrieval, most likely due to a too
harsh proteolysis in Pepsin, impaired the morphology as the nuclei
were almost totally digested, complicating the
identification of the BRISH signals. This
observation was not related to any specific system.
As discussed in run B9, the breast
ductal carcinoma no. 5 and 6, fixed for 48 and 72 hours
respectively, was much more challenging in order to perform an
optimal BRISH. This is most likely due to the prolonged fixation in
NBF compared to the 24 hours fixation time for the 4 other tissues
included in the multitissue block. However, an interpretation was possible
for the 16 laboratories assessed as optimal. For unexplained reasons
the tumour tissue no. 5 fixed for 48 h. was much more difficult to
stain than
the same tumour no. 6 fixed for 72 h.
The laboratories were requested to send in their own interpretation
on the stained sections. As regards amplification vs.
non-amplification 31 out of the 39 laboratories (80 %) assessed as
optimal or good and returning their own interpretation of the
tumours, interpreted and classified the tumours in concordance to
the HER-2/chr17 statuses generated in the reference
laboratories. 10 out of
the 31 laboratories interpreted the low level amplified tumour no. 4
(ratio of 2.3 - 2.6) as equivocal, which was accepted, as the
tumour according to general accepted guidelines would have been
re-evaluated. However, potentially the tumour could erroneously be
classified as non-amplified.
In the remaining 8 cases, 5 laboratories classified the low
amplified tumour, tissue specimen no. 4 as non-amplified and 3
laboratories classified the non-amplified tumour, tissue specimen
no. 3 as amplified.
In total 31 out of the 54 laboratories (57%) participating in this
assessment B10 and returning the slip obtained a sufficient mark
(good or optimal) and interpreted the result in concordance to the
reference HER-2 /chr17 statuses.
This was the fourth assessment of HER-2 BRISH in NordiQC. The pass
rate was the same as in the previous run (Table 2).
Table 2:
Proportion of sufficient results for HER-2 BRISH in four NordiQC runs
Conclusion
In this assessment an optimal demonstration of HER-2 BRISH could be
obtained by the two-colour HER-2 systems
INFORM™ HER-2 Dual SISH, Ventana and DuoCISH™, Dako. Also the
single-colour HER-2 system, INFORM™ HER-2 Dual SISH, Ventana could
be used to obtain an optimal demonstration.
For an optimal
performance the retrieval settings – HIER + proteolysis - must be
carefully balanced to provide an efficient sensitivity and preserved
morphology. Attention must also be addressed to the interpretation.
Only 57 % of the laboratories obtained a sufficient result together with a
correct interpretation.
|