The slide to be stained for CD79a comprised:
1. Appendix, 2. Tonsil, 3. Precursor-B-acute lymphatic leukaemia
(Pre-B-ALL), 4. B-chronic lymphatic leukaemia (B-CLL), 5.
Plasmacytoma.
All tissues were fixed in 10% neutral buffered formalin for 24-48 h.
Criteria for assessing a CD79a staining as optimal included:
- A strong, predominantly membranous
staining reaction of the mantle zone B-cells and at least a
moderate membranous staining reaction of the germinal centre
B-cells in the secondary follicles of the tonsil and the
appendix.
- A strong, predominantly cytoplasmic
staining reaction in the plasma cells and the late stage
activated germinal centre B-cells of the tonsil and the
appendix.
- A moderate to strong membranous staining
reaction of virtually all the neoplastic cells in the B-CLL.
- At least a weak to moderate predominantly membranous staining
reaction in the majority of the neoplastic cells in the
Pre-B-ALL.
- At least a weak to moderate cytoplasmic
staining reaction in the majority of the neoplastic cells in the plasmacytoma.
- No staining of any other cells.
141 laboratories participated in this
assessment. 75 % achieved a sufficient mark. In table 1 the
antibodies (Abs) used and marks are summarized.
Table 1.
Abs and
assessment marks
for CD79a, run 29
|
Concentrated Abs |
N |
Vendor |
Optimal |
Good |
Borderl. |
Poor |
Suff.1 |
Suff. OPS2 |
|
mAb3) clone
JCB117 |
84
3
1 |
Dako
NeoMarkers
Monosan |
43 |
32 |
9 |
4 |
85 % |
88 % |
|
mAb clone
HM57 |
8 |
Dako |
0 |
0 |
5 |
3 |
0 % |
- |
|
mAb clone
11D10 |
1 |
Novocastra/Leica |
0 |
0 |
0 |
1 |
- |
- |
|
rmAb clone
SP18 |
8
2
1 |
NeoMarkers
Spring Bioscience
Master Diagnostica |
7 |
3 |
1 |
0 |
91 % |
100 % |
|
Ready-To-Use Abs |
|
|
|
|
|
|
|
|
mAb clone
JCB117, IR621 |
13 |
Dako |
10 |
3 |
0 |
0 |
100 % |
100 % |
mAb clone
JCB117, 760-2630 |
12 |
Ventana |
0 |
4 |
6 |
2 |
33 % |
- |
mAb clone
JCB117, 179M-17 |
1 |
Cell Marque |
0 |
0 |
1 |
0 |
- |
- |
mAb clone
11E3, PA0192 |
2 |
Novocastra/Leica |
0 |
0 |
1 |
1 |
- |
- |
mAb clone
HM47/A9, PM067 |
1 |
Biocare |
0 |
0 |
1 |
0 |
- |
- |
|
rmAb clone
SP18, 760-4432 |
3 |
Ventana |
3 |
0 |
0 |
0 |
- |
- |
|
rmAb clone
SP18, ZA-0293 |
1 |
Unknown
(ZHI?) |
1 |
0 |
0 |
0 |
- |
- |
|
Total |
141 |
|
64 |
42 |
24 |
11 |
- |
- |
|
Proportion |
|
|
45 % |
30 % |
17 % |
8 % |
75 % |
- |
1) Proportion of sufficient stains
(optimal or good), 2) Proportion of sufficient stains with optimal
protocol settings only, see below.
3) mAb: mouse monoclonal
antibody, rmAb: rabbit monoclonal antibody, pAb: polyclonal
antibody.
Following central protocol parameters were
used to obtain an optimal staining:
Concentrated Abs
mAb clone JCB117: The protocols giving an optimal result were
all based on
HIER using either
Tris-EDTA/EGTA pH 9 (11/19)*, Target Retrieval Solution pH 9
(EnVision FLEX TRS high pH, Dako) (17/23), Target Retrieval Solution
pH 6.1 (S1699, Dako) (1/1), Cell Conditioning 1 (BenchMark, Ventana)
(3/22), Bond Epitope Retrieval Solution 2 (Bond, Leica) (7/10),
EDTA/EGTA pH8 (1/1) or Citrate pH 6 (3/7) as the retrieval buffer.
The mAb was typically diluted in the range of 1:25– 1:600 depending
on the total sensitivity of the protocol employed. Using these
protocol settings 65 out of 74 (88 %) laboratories produced a
sufficient staining (optimal or good).
* (number of optimal results/number of laboratories using this
buffer)
rmAb clone SP18: The protocols giving an optimal result were
based on HIER using Tris-EDTA/EGTA
pH 9 (2/3), Target Retrieval Solution pH 9 (EnVision FLEX TRS high
pH, Dako)(1/1), Target Retrieval Solution pH 6.1 (S1699, Dako) (1/1)
or Cell Conditioning 1 (BenchMark, Ventana) (3/3) as the retrieval
buffer. The mAb was typically diluted in the range of 1:100– 1:500
depending on the total sensitivity of the protocol employed. Using
these protocol settings all of 10 (100 %) laboratories produced a
sufficient staining.
Ready-To-Use Abs
mAb clone JCB117 (prod. no IR621, Dako): The protocols giving
an optimal result were all based on HIER in PT-Link using Target
Retrieval Solution pH 9 (EnVision FLEX TRS high pH), an incubation
time of 20 min in the primary Ab and EnVision Flex (K8000) or Flex+
(K8002) as the detection system. Using these protocol settings all of 12 (100 %) laboratories produced a sufficient staining.
rmAb clone SP18 (prod. no. 760-4432, Ventana), The protocols
giving an optimal result were based on HIER using standard Cell
Conditioning 1, an incubation time of 16-48 min in the primary Ab
and UltraView (760-500) as the detection system. Using these
protocol settings all of 3 laboratories produced an optimal staining.
rmAb clone SP18 (ZA-0293, Unknown -
ZHI?), the protocol giving an
optimal result was bases on HIER in Citrate pH 6 in a pressure
cooker, an incubation time of 60 min in the primary Ab and a polymer
based detection system (Jinqiao Zhongshan, D11-110).
- - -
The most frequent causes of insufficient staining were:
- Less successful primary antibody (e.g., all of 8 stains based
the mAb clone HM57 were insufficient)
- Less successful RTU format (e.g., 9 out of 13 stains based on the mAb clone JCB117 (Ventana/Cell Marque) were insufficient)
- Too low concentration of the primary antibody
- Insufficient HIER (e.g., too short efficient heating time).
In this assessment and in concordance to the previous assessments of
CD79a, the prevalent feature of an insufficient staining was a too
weak or false negative reaction of cells expected to be
demonstrated. The majority of the laboratories were able to
demonstrate CD79a in the normal plasma cells, the mantle zone
B-cells and the neoplastic cells of the B-CLL. However, the
demonstration of CD79a in both the Pre-B-ALL and the plasmacytoma
was much more difficult and only obtained when using correctly
calibrated protocols based on mAb clone JCB117 or rmAb clone
SP18 with an efficient HIER. The mAb clone HM57 has a too low
affinity for CD79a, as also found in this assessment, where all of 8 protocols based on this clone gave
an insufficient result. At the same time HM57 gave a strong
cross-reaction with the smooth muscle cells and appendiceal
enterocytes. Clone HM57, Dako, has now been withdrawn from the
market. The mAb clone HM47/A9
(Biocare) gave the same poor staining pattern as clone HM57. As observed
in run 21, 2007, the mAb clone 11E3 (RTU,
Novocastra/Leica) gave a distinct reaction in
plasma cells and the mantle zone B-cells, but only a faint staining
of the
germinal centre cells and virtually negative staining of the plasmacytoma and
Pre-B-ALL.
Table 2:
Performance of the four most commonly used Abs in three CD79a runs
| |
Run 6
2002 |
Run 21 2007 |
Run 29 2010 |
Total |
|
|
Protocols |
Sufficient |
Protocols |
Sufficient |
Protocols |
Sufficient |
Protocols |
Sufficient |
|
mAb clone JCB117 |
48 |
43 |
97 |
80 |
124 |
92 |
269 |
215 (80 %) |
|
mAb clone HM57 |
3 |
0 |
6 |
0 |
8 |
0 |
17 |
0 (0 %) |
|
mAb clone 11E3 |
0 |
0 |
2 |
0 |
2 |
0 |
4 |
4 (0 %) |
|
rmAb clone SP18 |
0 |
0 |
6 |
6 |
11 |
10 |
17 |
16 (94 %) |
The staining results for
the mAb clone JCB117 were influenced by the stainer
platform: When the clone was used on, e.g., the Dako Autostainer platform, 36 out of 39 of the protocols
(92%) gave a sufficient result (of which 30 were optimal - 77 %). In
contrast, on the Ventana BenchMark
instrument,
only 17 out of 25 protocols (68%) gave a sufficient staining, of
which 3 (12 %)
resulted in an optimal staining, the latter all based on a high Ab
concentration (1:25 – 1:100).
The Dako RTU format of
the mAb clone JCB117 was very robust
as all of 13 stains based on this IHC platform were assessed as
sufficient (10 of them optimal). In contrast only 4 of 13 stains
based on the same clone and the RTU system from Ventana were
assessed as sufficient (and none were assessed as optimal). However,
the RTU format of the rmAb clone SP18, Ventana gave an
optimal result in all of 3 laboratories using the Ventana IHC platform.
This was the 3rd assessment of CD79a in NordiQC. A relatively constant
proportion of sufficient results have been seen in the three runs (table 3).
Table 3:
Proportion of sufficient results for CD79a in the three NordiQC runs
performed
Tonsil and appendix are reliable positive
controls: The germinal centre B-cells must show at least a moderate
and distinct membranous reaction. If only the mantle zone B-cells is
distinctively demonstrated, the protocol is too insensitive to
detect CD79a in neoplasias with a low CD79a expression such as plasmacytoma and Pre-B-ALL.
Conclusion
The mAb clone JCB117 and the rmAb SP18 are both recommendable Abs
for the demonstration of CD79a. HIER is mandatory to obtain an
optimal result. Concentration of the primary Ab should be carefully
calibrated. Tonsil and appendix are appropriate controls: The
germinal centre B-cells must show at least a moderate and distinct
membranous staining reaction, while no staining is to be seen in the
smooth muscle cells. |