Home  ■  Protocols

Recommended protocol Run 25 2009
Aliso Viejo, USA

CD30 (1G12)

Protocol provided by

 
  laboratory Clarient Diagnostic Services
  hospital/institution Private laboratory, Aliso Viejo, USA
  e-mail address eluckey@clarientinc.com
Primary antibody  
  clone 1G12
  producer Novocastra
  product no., lot no. NCL-L-CD30, L115336
  dilution 1:40
  diluent buffer and additive(s) Ab diluent, BOND
  incubation time/temp. 30 min./RT
Epitope retrieval, HIER  
  device BOND-Max
  buffer type, pH Bond Epitope Retrieval Solution 2
  warm-up/heating max/resting time 1 min./30 min./1 min.
  maximum heating temp. 100ºC
Visualization system  
  method Polymer conjugate
  producer, product no. Leica Microsystems, DS9800
  incubation time/temp. 16 min./RT
Chromogen  
  type DAB
  producer, product no. Leica Microsystems, DS9800
  incubation time/temp. 10 min./RT
  enhancement, type NS
Immunostainer  
  type BOND-Max, Leica Microsystems
     
Abbreviations  
  HIER: heat induced epitope retrieval, MWO: microwave oven, NS: not stated, RT: room temperature
Disclaimer  
  NordiQC makes every attempt to provide accurate and up-to-date information, yet NordiQC does not make any claim or warranty regarding the accuracy of the provided information nor does it represent that the contents of the web site and protocols reflect the most recent developments in immunohistochemistry at any point in time.
 
SN/AS

Last update 06-04-2009