The
slide to be stained for
epithelial cell-cell adhesion molecule
(Ep-CAM) comprised:
1. Colon adenocarcinoma, 2. Kidney, 3. Appendix, 4. Clear cell renal cell
carcinoma, 5. Chromophobe renal cell
carcinoma, 6. Ileal carcinoid.
All tissues were fixed in 10% neutral buffered formalin.
Criteria for assessing a membranous Ep-CAM
staining as optimal included:
- A strong, distinct, predominantly
membranous staining of virtually all the columnar epithelial
cells in the appendix.
- A moderate to strong predominantly
membranous staining of the epithelial cells of the renal
collecting tubules, and of scattered epithelial cells lining the
Bowman capsule.
- A strong, distinct, predominantly
membranous staining of virtually all the neoplastic cells of the
colon adenocarcinoma and the ileal carcinoid.
- A moderate to strong predominantly
membranous staining of the majority of the neoplastic cells of
the two renal cell carcinomas.
83 laboratories submitted stains. 5
laboratories used an inappropriate Ab (such as low molecular weight
cytokeratin). At the assessment of 78 laboratories, 29 achieved
optimal marks (37 %), 20 good (26 %), 13 borderline (17 %) and 16
poor marks (20 %).
The following Abs were used:
mAb clone Ber-EP4 (Dako, n=67; NeoMarkers/Thermo Scientific,
n=4)
mAb clone MOC-31 (Dako, n=4; Novocastra, n=1)
mAb clone KS1/4 (BD Pharmingen, n=1)
mAb clone VU-1D9 (Euro-Diagnostica, n=1)
Optimal staining for Ep-CAM in this assessment
was obtained with the mAb clones Ber-EP4 (27 out of 71),
MOC-31 (1 out of 5) and VU-1D9 (1 out of 1).
With mAb clone Ber-EP4 the optimal
protocols were based on either heat induced epitope retrieval (HIER)
or enzymatic pre-treatment.
HIER was the most successful epitope retrieval method (Table 1).
Table 1.
The proportion of sufficient and optimal results with HIER and
proteolytic
pre-treatment, respectively.
| |
HIER |
Proteolysis |
| |
Sufficient |
Optimal |
Sufficient |
Optimal |
|
mAb
Ber-EP4 |
76% (37/49) |
51% (25/49) |
36% (8/22) |
9% (2/22) |
The most robust HIER buffer was Target Retrieval Solution low pH 6.1
(Dako, TRS, S1699/1700) as 32 out of 33 stains (97%) gave a sufficient
result (out of which 22 [67%] were optimal). Citrate pH
6.0 gave a sufficient result in 3 out of 8 results (of which 2 [25%]
were optimal).
HIER in an alkaline buffer such as Tris-EDTA/EGTA pH 9.0 or Cell
Conditioning 1 (Ventana) gave no sufficient
results out of 8.
Using HIER, the mAb was typically used in the range of 1:25 – 1:500
depending on the total sensitivity of the protocol employed or as a
Ready-To-Use (RTU) Ab.
The two optimal results with proteolytic pre-treatment were based
on either 0.1 % Trypsin or Bond™ Enzyme (Bond, Leica Microsystems).
Using proteolytic pre-treatment, the mAb was used in the range of
1:25 - 1:100.
With mAb clone MOC-31 the optimal
protocol was based on HIER in Target Retrieval Solution pH 6.1
(Dako, TRS) and a dilution of 1:25 of the primary Ab.
With mAb clone VU-1D9 the optimal
protocol was based on enzyme pre-treatment in Protease Type XIV 0.03
and a dilution of 1:100 of the primary Ab.
The most frequent causes of insufficient
staining were:
- Inappropriate HIER buffer (alkaline)
- Insufficient proteolytic pre-treatment
- Excessive proteolytic pre-treatment
- Too low concentration of the primary Ab.
The majority of the
laboratories were able to demonstrate Ep-CAM in the columnar
epithelial cells of the appendix and the neoplastic cells of the
colon adenocarcinoma and the carcinoid, whereas the prevalent
feature of the insufficient staining was a too weak or false
negative staining of the two renal cell carcinomas, particular
the clear cell type. When proteolysis was used, it generally gave
either a too weak reaction in the renal cell carcinomas due to
insufficient pre-treatment or a false negative reaction due to excessive
proteolysis where the cytoplasm and cell membranes were partly or
totally eroded.
In accordance with the previous assessments for Ep-CAM normal
kidney was an applicable control for Ep-CAM, as the ability to demonstrate Ep-CAM
in the collecting tubules and epithelial cells lining the Bowman’s
capsule was characteristic for laboratories obtaining an optimal
mark.
Table 2. Pass rate for Ep-CAM
in three consecutive assessments.
Table 2 shows the proportion of sufficient
results in the two previous runs and the current run. The reason for the pass rate being lower in
the two latest runs compared to the first run is probably a more
challenging material. The specifically tailored recommendations
given to the laboratories obtaining an insufficient mark in run 17
did have a positive impact: 24 laboratories which obtained an
insufficient result in run 17, submitted a new Ep-CAM stain in run
23. 10 of these followed the recommendation, of which 8 improved to
good or optimal (80 %). 10 laboratories did not follow the
recommendation, and only 2 of these (20 %) obtained a sufficient
staining in run 23. 4 laboratories changed their entire IHC system.
Of these 3 obtained a sufficient mark.
The tailored recommendations for Ep-CAM to the laboratories are
relative problematic for NordiQC. The most robust method appears to be
based on HIER in Target Retrieval Solution low pH 6.1, (Dako), which
is adaptable to all open IHC systems. For "closed" staining
systems with on-board HIER using the reagents developed and
optimized for the respective IHC-system (e.g., Bond™, Leica
Microsystems and BenchMark XT, Ventana), NordiQC has not been able
to identify any robust protocol in the three Ep-CAM assessments. For these platforms, only proteolysis has
been used to obtain an optimal result, but as shown in table 1, the
success rate is very low. NordiQC will collaborate with the
respective companies to develop useful protocols for Ep-CAM.
Conclusion
The mAb clones Ber-EP4, MOC31 and VU-1D9 are
all useful Abs for Ep-CAM. HIER in Target Retrieval Solution low
pH 6.1 (Dako) is recommended for optimal performance for Ber-EP4 and
MOC31. Kidney is an applicable control: At least a moderate membranous staining should be seen in the
epithelial cells of the renal collecting tubules and scattered
epithelial cells lining the Bowman capsule. |