The
slide to be stained for Chromogranin A
(CGA)
comprised:
1. Brain, 2. Appendix, 3. Colon adenocarcinoma, 4. Pancreas (from
patient with blood group A),
5. Pancreas (from patiet with blood group 0), 6-7. Endocrine lung carcinomas, 8. Medullary thyroid carcinoma.
All tissues were fixed in 10% neutral buffered formalin.
Criteria for assessing a CGA staining as optimal included:
- A strong and distinct cytoplasmic reaction of the normal
neuroendocrine cells in the appendiceal mucosa and in the islet cell
in the pancreas*
- A moderate to strong distinct cytoplasmic
reaction in the normal ganglion cells and axons in the appendiceal
Aurbach’s plexus as well as the cortical neurons in the brain
- A strong distinct cytoplasmic reaction of
the medullary thyroid carcinoma
- A moderate to strong distinct cytoplasmic
reaction of the two endocrine lungcarcinomas
- A negative reaction of the neoplastic cells in the colon
adenocarcinoma.
* A weak background reaction
(due to diffusion of the
antigen/chromogen and focal autolysis in the two
pancreas specimens) was inevitable.
117 laboratories submitted stains. At the assessment 34 achieved
optimal marks (29 %), 37 good (32 %), 25 borderline (21 %) and 21
poor marks (18 %).
The following Abs were used:
mAb clone LK2H10 (Ventana, n=11; NeoMarkers, n=8; BioGenex,
n=5; Linaris, n=4; Chemicon, n=1; Euro-Diagnostica, n=1; Novocastra, n=1)
mAb clone LK2H10+PHE5 (NeoMarkers, n=5; BioCare Medical, n=1)
mAb clone DAK-A3 (Dako, n=16)
mAb clone 5H7 (Novocastra, n=2)
rmAb clone SP12 (NeoMarkers, n=4)
pAb A0430 (Dako, n=57)
pAb N1535 (Dako, n=1)
pAb 18-0094 (Zymed, n=1)
Optimal staining for CGA in this assessment was obtained with the
mAb clone LK2H10 (8 out of 30), the mAb clone LK2H10+PHE5 (2 out of
6) and the pAb A0430 (24 out of 56).
LK2H10: the protocols giving an optimal result
were all based on heat induced epitope retrieval (HIER) using either
Tris-EDTA/EGTA pH 9 (3/10)*, Cell Conditioning1 (BenchMark, Ventana; 4/10) or Citrate pH 6 (1/4). The mAb was typically diluted in the
range of 1:500 – 1:1.000 depending on the total sensitivity of the
protocol employed or as a Ready-To-Use antibody. Using these
protocol settings 17 out of 20 (85 %) laboratories produced a
sufficient staining (optimal or good).
*number of optimal results/number of laboratories using this
buffer
LK2H10+PHE5: the protocols giving an optimal result were all
based on HIER using Tris-EDTA/EGTA pH 9 (2/4). The mAb was diluted in the range of
1:200 – 1:5.000 depending on the total sensitivity of the protocol
employed. Using these protocol settings 4 out of 4 (100 %)
laboratories produced a sufficient staining.
A0430: the protocols giving an optimal result were all
based on HIER using Tris-EDTA/EGTA pH 9 (20/37), Cell Conditioning1
(BenchMark, Ventana) (2/7), Bond Epitope Retrieval Solution 1 (Bond,
Vision Biosystems) (1/4) or Target Retrieval Solution pH 9 (FLEX TRS
high pH, Dako) (1/3). The pAb was typically diluted in the range of
1:250 – 1:4.000 depending on the total sensitivity of the protocol
employed. Using these protocol settings 33 out of 39 (85 %)
laboratories produced a sufficient staining.
The most frequent causes of insufficient staining were:
- Less successful primary antibody
- Too low concentration of the primary antibody
- Omission of HIER
In this assessment and in concordance with the observations in
previous CGA assessments almost all laboratories were able to
demonstrate CGA in the normal neuroendocrine cells in e.g. pancreas,
whereas the prevalent feature of the insufficient staining was a too
weak or false negative staining in the neurons and axons as well as
the three neuroendocrine carcinomas.
A too weak or false negative staining was seen in 98 % of the
insufficient results (45 out of 46).
In all the assessments for CGA, appendix have shown to be a reliable control,
as a moderate to strong staining of the the Aurbach’s plexus
(neurons and axons) being the critical stain quality indicator for a
proper immunohistochemical
demonstration of CGA in the tumours.
The choice of Ab and epitope retrieval are central parameters for a successful demonstration of CGA.
In Table 1, the overall pass rates are compared
for the most widely used Abs in the latest three assessments:
|
Table 1 |
Run 13 2005 |
Run 18 2006 |
Run 22 2008 |
Total |
| |
Protocols |
Sufficient |
Protocols |
Sufficient |
Protocols |
Sufficient |
Protocols |
Sufficient |
| mAb clone
LK2H10 |
22 |
14 |
19 |
14 |
31 |
25 |
72 |
53 (74%) |
| mAb clone
LK2H10+PHE5 |
3 |
3 |
5 |
3 |
6 |
4 |
14 |
10 (71%) |
| mAb clone
DAK-A3 |
15 |
2 |
11 |
0 |
16 |
0 |
52 |
2 (4%) |
| pAb A0430 |
44 |
37 |
53 |
42 |
57 |
39 |
154 |
118 (77%) |
Thus, in 3 successive assessments only 4%
of the slides stained with mAb clone DAK-A3 showed a sufficient result
(compared to 71-77% for the 3 other
Abs). In accordance with this, tests in the NordiQC laboratory have
shown that it is not possible to obtain satisfactory results with
DAK-A3. In spite of the clear demonstration of its insensitivity in
the previous runs, 16 labs used DAK-A3 in the current run. NordiQC
will again urge the producer to withdraw the clone from the market
an inform the costumers about the problem.
Omission of HIER is another important protocol parameter. 7 out of 8 protocols using
mAb clone LK2H10 or pAb A0430 without HIER in the current run gave insufficient results.
This was the 4th NordiQC assessment of CGA. The
proportion of sufficient results in the four runs are shown in Table
2.
|
Table 2 |
Run 9
2003 |
Run 13
2005 |
Run 18
2006 |
Run 22 2008 |
|
Participants, n |
77 |
88 |
94 |
117 |
|
Sufficient
results |
49% |
64% |
70% |
61% |
Major reasons for the unsatisfactory low
proportion of sufficient staining results are the use of the low
affinity mAb DAK-A3 and omission of HIER.
68 laboratories participating in
all 4 runs for CGA showed an increase in the proportion of sufficient
results from 41 % (run 9) to 74 % (run 22). 89 laboratories
participated in both run 18 and run
22. For these, the proportion of sufficient results were 71% and
73%, respectively. The decline in the over all proportion of
sufficient stains from run 18 to run 22 is consequently due to a
large number of new participants submitting CGA stains for the first
time. Conclusion
The mAb clones LK2H10 and LK2H10 + PHE5 and the pAb A0430 are useful
Abs for the demonstration of CGA. HIER, preferentially in an alkaline
buffer, and a proper calibration of Ab titer are mandatory to obtain
an optimal result. Normal appendix is a recommendable control tissue: the ganglion cells and axons in the Aurbach’s plexus
must show an
at least moderate, distinct granular reaction, while the surrounding
muscle cells should be unstained. |