7
laboratories participated. In one gastrointestinal stromal tumour
(GIST), good or moderate staining was obtained in 6 laboratories.
However, in another GIST, good staining was obtained with one
protocol only, while three protocols gave acceptable staining.
Dako's polyclonal antibody appeared most reliable. Less optimal
stainings appeared when the antibody was too diluted. Another
cause for poor stainings appeared to be insufficient demasking
(insufficient heating time, buffer pH too low), see Epitope retrieval.
|